Welcome to Mediso

Please select your region:

North America
Europe

Please select your language:

Go

AnyScan® TRIO SPECT/CT TheraMAX*

Theranostic and Diagnostic Imaging with MAXimum Performance

Explore new horizons with the AnyScan® TRIO SPECT/CT TheraMAX* featuring extreme sensitivity with uncompromised image quality even at high energies. Transform your clinical routine with ultra-fast quantitative total-body scans in the era of theranostics. A versatile SPECT/CT system designed with the endeavour of serving the clinical needs of Nuclear Medicine and Molecular Imaging.

*TheraMAX is the configuration name of MAX-123/15.9 detector

Features & Benefits

MAXimized Performance for Theranostics

  • Theranostic applications with isotope energies up to 640 keV provided by 15.9 mm NaI crystal 
  • Detector sensitivity gain of 30% for 177Lu, 80% for both 131I and 225Ac when compared to 9.5 mm NaI crystal based detector 
  • Intrinsic resolution of 3.3 mm (FWHM) maintained by high-density sensor arrangement (123 PMTs per detector head)

Ultra-fast Quantitative Total-Body SPECT/CT

  • Extreme sensitivity of 1620 kcps/(MBq/cm3) for 99mTc measured as Total System Response** delivered by the 15.9 mm crystal thickness 
  • Triple-NaI-Detector: patient surrounded completely in 360° with three large surface detectors (up to 5x larger than 12-detector CZT systems)
  • The continuous detector movements provide seamless SPECT acquisition
  • Long Axial-FOV of 40 cm enables 2 meters SPECT/CT imaging in only five bed positions and ultra-fast quantitative total-body scan with acquisition time less than 10 minutes

**NEMA NU1 2018 parameter calculated as Average Volume Sensitivity per Axial Centimeter × Axial FOV

From General Purpose to Organ Focus Imaging

  • The TheraMAX* is a versatile SPECT/CT system offering five imaging modes using Transformable™ Gantry and simple collimator exchange procedures
  • Total-Body mode unveils ultra-fast quantitative total-body SPECT/CT by all three detectors surrounding the patient
  • Brain focus imaging is possible with the dedicated MPH-Brain collimator for superior image quality DAT SPECT and Brain Perfusion SPECT
  • Cardiac focus imaging with the MPH-Cardiac collimator delivers extreme tomographic sensitivity enabling fast myocardial perfusion or amyloid scans and even dynamic SPECT acquisition
  • Dual-detector modes provide real planar and SPECT imaging options

MAXimized Image Quality and Quantitative Accuracy

  • Our signature parallel-hole collimators are available for low-, medium-, high- and ultra-high energies.  
  • Novel multi-pinhole technology provides focused imaging for brain and cardiac applications with PET-like image quality and unveils dynamic SPECT applications.  
  • Tera-Tomo™ 3D SPECT-Q is the ultimate quantitative image reconstruction solution, featuring Monte Carlo based physical modelling of particle-level gamma photon interactions running on high-performance GPUs. 

Triple-Performance CT

  • High power CT with up to 440 mA in diagnostic mode, 0.5 sec rotation speed, 0.625 mm resolution while offering on-the-fly dose modulation
  • 16 slice diagnostic CT with 32-slice reconstruction offering three performance levels: 

Standard Low Dose protocols for diagnostic CT applications 

Ultra-Low Dose CT with sub-mSv effective dose for attenuation correction and localization 

AI-Powered Synthetic CT (SyCT)** with zero effective CT dose for quantitative SPECT attenuation correction and anatomical localization in 99mTc bone imaging    

  • Iterative Image Reconstruction with Tera-Tomo™ 3D CT-IR and advanced corrections for Motion- and Metal Artefact Reduction
  • Further improvement is possible with AI-based CT image quality enhancement solution.

***For research purpose only 

SyCT***– the AI-Powered Synthetic CT

Experience the innovation of zero-dose CT imaging for quantitative SPECT analysis with our AI-Powered Synthetic CT (SyCT) solution:

  • Free from the effective dose of traditional CT scans
  • Accurate SPECT attenuation correction for absolute quantification
  • Free from artifacts associated with conventional CT imaging
  • No motion and registration mismatches between SPECT and CT
  • Anatomical localization
  • Enhanced workflow efficiency: no need for dual-modality imaging
  • Applicable for 99mTc bone scans, including total-body imaging

First Global Installation of the AnyScan® TRIO SPECT/CT, TheraMAX

“The AnyScan® TRIO SPECT/CT, TheraMAX will boost our diagnostic and theranostic workflow and targeted radionuclide therapy, leading to better patient care and it also opens new possibilities in clinical research.” Said Professor Dr. Dirk Hellwig, the head of the Department of Nuclear Medicine at University Hospital Regensburg, Germany.

Applications

Publications

2024.03.19.

Dynamic multi-pinhole collimated brain SPECT of Parkinson’s disease by [123I]FP-CIT: a feasibility study...

View publication
2024.01.18.

Dopamine Transporter SPECT with 12-Minute Scan Duration Using Multiple-Pinhole Collimators

View publication

Images

Specification

SPECT
CT
Detector Crystal
NaI(Tl) 15.9 mm
Bore Size
75 cm
Scan range
200 cm
FOV
545×401 mm
Energy range
35 - 640 keV
Ultra-fast Quantitative Total-Body SPECT/CT
in less than 10 min
Head positions
120°, 180°, 90° or dedicated cardiac position (75°)
Multi-pinhole collimators
Brain, Cardiac, other organ specific
Low Energy Collimators
Low Energy High Resolution & High Sensitivity (LEHR-HS),                            Low Energy Ultra High Resolution & High Sensitivity (LEUHR-HS)
Medium to High Energy Collimators
Medium Energy High Resolution & High Sensitivity (MEHR-HS),                          Medium High Energy General Purpose (M-HEGP)
High to Ultra-High Energy Collimators
High Ultra-High Energy General Purpose (H-UHEGP)
Rotation speed
0.5 - 2 sec
Bore size
70 cm
Tube current
10 - 440 mA
Anode heat capacity
7.5 MHU
Number of acquired slices
16
Diagnostic CT
Yes
Resolution
Sub-millimeter: 0.625 mm

How can we help you?

Don't hesitate to contact us for technical information or to find out more about our products and services.

Get in touch